2019, vol. 74, iss. 1, pp. 159-163
Uncertainty evaluation of the results of electrical insulation tester calibration and its application for testing
Beograd

emailnvucijak@gmail.com
Keywords: electrical insulation resistance; calibration; testing; uncertainty of measurement
Abstract
In this paper evaluation of the uncertainty of electrical insulation tester calibration and evaluation of the uncertainty of electrical insulation testing results, are presented in details. The uncertainty of measurements is evaluated according guides whose abbreviated marks are: GUM and EA - 4/02 M:2013. Traceability of standards used in calibration and traceability of measuring instruments used in testing, is obliged. The calibration of measuring instrument is performed at reference conditions by accredited laboratory for calibration. When using calibrated measuring instrument for outdoor measurement then the working conditions may be different from the reference conditions, and expanded uncertainty can have larger value.
References
*** (2017) SRPS ISO/IEC 17025:2017 opšti zahtevi za kompetentnost laboratorija za ispitivanje i laboratorija za etaloniranje. Institut za standardizaciju Srbije, decembar
*** (2013) EA - 4/02 M:2013 evaluation of the uncertainty of measurement: European co-operation for accreditation. www.european-accreditation.org
*** (2010) Pravilnik o vrstama merila za koja je obavezno overavanje i vremenskim intervalima njihovog periodičnog overavanja. Sl. Glasnik RS, br. 49
*** (2003) EA-4/16 G (rev.00):2003: EA guidelines on the expression of uncertainty in quantitative testing. European co-operation for Accreditation, www.european-accreditation.org
*** (2002) ILAC-G17:2002 introducing the concept of uncertainty of measurement in testing in association with the application of the standard ISO/IEC 17025. https://ilac.org
*** (2008) JCGM 200:2008 International vocabulary of metrology: Basic and general concepts and associated terms. www.BIPM.org
*** (2008) JCGM 100:2008: (GUM 1995 with minor corrections) evaluation of measurement data: Guide to the expression of uncertainty in measurement. www.BIPM.org
*** (2008) JCGM 101:2008: Evaluation of measurement data - supplement 1 to the guide to the expression of uncertainty in measurement: Propagation of distributions using a Monte Carlo method. Joint Committee for Guides in Metrology, www.-BIPM.org
Vučijak, N.M., Pejović, P.V. (2014) Dva postupka procene merne nesigurnosti rezultata merenja fazne razlike metodom tri voltmetra. Tehnika, vol. 69, br. 3, str. 549-554
 