Metrics

  • citations in SCIndeks: [1]
  • citations in CrossRef:[1]
  • citations in Google Scholar:[]
  • visits in previous 30 days:6
  • full-text downloads in 30 days:3

Contents

article: 2 from 2  
Back back to result list
2014, vol. 69, iss. 3, pp. 549-554
Two procedures of evaluation of uncertainty for results of phase difference measurement using three voltmeter method
aBeograd
bUniversity of Belgrade, Faculty of Electrical Engineering
Keywords: phase difference; three voltmeter method; uncertainty of measurement; GUM; Supplement1 to GUM
Abstract
Phase difference measurement between two sinusoidal electric voltages using three voltmeter method is presented. Uncertainty of phase difference measurement is evaluated in two methods: according to Guide for expression of uncertainty in measurement (GUM) and according to Supplement 1 to GUM - Propagation of distributions using a Monte Carlo method. Those results for measurement uncertainty are compared.
References
*** (2008) ISO IEC GUIDE 98-3/Suppl.1: Uncertainty of measurement Part 3: Guide to the expression of uncertainty in measurement (GUM:1995) Supplement 1: Propagation of distributions using a Monte Carlo method
*** (2008) GUIDE OIML G 1-100: Evaluation of measurement data: Guide to the expression of uncertainty in measurement. OIML
*** (2008) ISO IEC GUIDE 98-3: Uncertainty of measurement: Guide to the expression of uncertainty in measurement. Part 3, (GUM:1995)
*** (2008) JCGM 200:2008 International vocabulary of metrology: Basic and general concepts and associated terms. VIM, 3rd edition
*** (2000) 3458A Multimeter: Users guide. Agilent
*** Calibration and measurement capabilities: CMC. http://kcdb.bipm.org/appendixc/EM/DE/EM_DE
*** (1995) Guide to the expression of uncertainty in measurement. Switzerland: ISO, BIPM, IEC, IFCC, ISO, IUPAC, IUPAP and OIML
Callegaro, L., D`Elia, V. (2001) Automated system for inductance realization traceable to AC resistance with a three-voltmeter method. IEEE Transactions on Instrumentation and Measurement, 50(6): 1630-1633
Clarke, K.K., Hess, D.T. (1990) Phase measurement, traceability, and verification theory and practice. IEEE Transactions on Instrumentation and Measurement, 39(1): 52-55
Joint Committee for Guides in Metrology (2006) Evaluation of measurement data, supplement 1 to the 'Guide to the expression of uncertainty in measurement': Propagation of distributions using a Monte Carlo method. Nacrt
Joint Committee for Guides in Metrology (2006) Evaluation of measurement data, supplement 1 to the 'Guide to the expression of uncertainty in measurement': Propagation of distributions using a Monte Carlo method. Nacrt
Mohns, E., Kahmann, M. (2007) Heterodyne Measurement System (HMS) for Determining Phase Angles. IEEE Transactions on Instrumentation and Measurement, 56(2): 505-508
Ramm, G., Moser, H., Braun, A. (1999) A new scheme for generating and measuring active, reactive, and apparent power at power frequencies with uncertainties of 2.5×10/sup -6/. IEEE Transactions on Instrumentation and Measurement, 48(2): 422-426
Svenson, S. (1999) Power measurement techniques for non-sinusoidal conditions. Goteborg, Sweden: Chalmers university of technology, Doctoral thesis for the degree of PhD
Webster, J., ed. (1999) The measurement, instrumentation and sensors handbook. CRC Press, poglavlje 41
 

About

article language: Serbian
document type: Professional Paper
DOI: 10.5937/tehnika1403549V
published in SCIndeks: 11/06/2015

Related records